Skip to content

Laboratory Quality Control Protocols

QC Framework

SENO operates a multi-level QC system integrated into every testing process. This system is designed to ensure that every test result released from the Zhangjiakou laboratory meets the highest standards of accuracy, reliability, and reproducibility. The QC framework is aligned with ISO 9001:2015 (quality management) and ISO 13485:2016 (medical device QMS) requirements and is subject to both internal and external audit.

Per-Batch Controls

Every PCR or qPCR run includes the following controls to validate assay performance and detect potential issues:

Control TypeFrequencyPurposeExpected ResultAction on Failure
Positive controlEvery runConfirm assay functionStrong amplification (expected Ct range)Investigate reagents, repeat run
Negative controlEvery runDetect contaminationNo amplification (No Ct)Investigate source, quarantine affected reagents
No-template controlEvery runVerify reagent purityNo amplification (No Ct)Replace reagents, repeat run
Internal amplification controlEvery sampleConfirm DNA quality and absence of inhibitionAmplification within expected rangeRe-extract DNA, dilute if inhibition suspected
Extraction blankEvery extraction batchMonitor extraction reagent contaminationNo amplification (No Ct)Replace extraction reagents, re-extract batch

Control Pass/Fail Decision Matrix

Control ResultInterpretationAction
Positive Ct valid, Negative NTC no Ct, IAC validRun passedRelease results
Positive Ct out of rangeAssay degradationRepeat with fresh reagents
Negative Ct detected (positive NTC)ContaminationInvestigate source, repeat all samples
IAC failed (no Ct) in healthy controlInhibitionRe-extract or dilute sample
Extraction blank positiveReagent contaminationReplace extraction reagents

Daily QC

Daily QC checks ensure that laboratory equipment and environment remain within validated parameters:

Check ItemParameterMethodAction on Failure
PCR machine temperatureBlock uniformity ± 0.5°CThermal validation plateService engineer call
Freezer temperature-20°C ± 3°C / -80°C ± 5°CDigital data logger + alarmTransfer samples, service freezer
Refrigerator temperature4°C ± 2°CDigital data logger + alarmTransfer reagents, service unit
Pipette accuracyWithin spec per volumeGravimetric check (water weight)Recalibrate, quarantine affected results
Room temperature20-25°CWall-mounted thermometerAdjust HVAC
Laboratory environmentSurface swabs, air samplesContact plates, settling platesIf positive: clean, identify source, retest
Reagent performanceControl sample amplificationRun positive control with new lotInvestigate lot, contact supplier

Periodic QC

ActivityFrequencyResponsibleDocumentation
Instrument calibrationQuarterly (or per manufacturer)External calibration serviceCalibration certificate
Proficiency test participationSemi-annual (annual minimum)QC ManagerPT report + evaluation
SOP review and updateAnnualQC Manager + Lab ManagerSOP revision log
Management reviewSemi-annual (quarterly minimum)CEO + QC Manager + Lab ManagerReview meeting minutes
Internal auditQuarterlyTrained internal auditorAudit report + CAPA
Reagent lot-to-lot validationEach new lotQC technicianLot validation report
Method re-validationAnnual or after significant changeR&D + QCValidation report
Equipment maintenancePer manufacturer scheduleLab ManagerMaintenance log
Staff competency assessmentAnnualLab ManagerCompetency checklist
Data integrity auditSemi-annualQC ManagerAudit trail review

Sample Retesting Protocol

SENO retests approximately 20% of samples for QC purposes. This robust retesting program ensures the accuracy and reliability of every result:

Categories Requiring Retesting

CategoryCriteriaRetesting Scope
First-time resultsAny species, any test type — first submissionFull repeat from DNA extraction
Ambiguous resultsWeak amplification, unclear band patternFull repeat from DNA extraction
Inconclusive resultsIAC failure, conflicting replicatesRe-extract + repeat
Client verification requestsClient asks for result confirmationFull repeat (no additional charge)
Statistical random selection~5% of routine resultsFull repeat from DNA extraction
Post-QC failure re-runsPrevious run failed controlsRepeat extraction + assay

Retesting Workflow

  1. Identification — Result flagged as requiring retest (automated in LIMS or manual by technician)
  2. Assignment — Retest assigned to different technician (ensures independence)
  3. Re-extraction — New DNA extraction from original sample (if sufficient material)
  4. Re-amplification — Fresh PCR/qPCR run with new controls
  5. Comparison — Original vs. retest results compared
  6. Reconciliation — If results match: release. If results differ: investigate and triplicate repeat
  7. Documentation — All retesting logged in LIMS with full traceability

Result Release Criteria

Results are released only when all of the following conditions are met:

CriterionVerification MethodDocumentation
All controls produce expected resultsVisual check of amplification curves / gel imagesRun log entry
Internal amplification control is validIAC Ct within expected range for each sampleSample-level QC flag
Result interpretation is unambiguousTwo-band pattern (female) or single-band (male) clearly visibleTechnician interpretation
Reviewed by second qualified technicianIndependent review of raw data and interpretationSecond review signature in LIMS
Retesting completed (if required)Retest results match original or discrepancy resolvedRetest log entry
No outstanding non-conformancesCAPA status checked for affected assaysLIMS workflow gate

Equipment Calibration Schedule

EquipmentCalibration FrequencyMethodAcceptance Criteria
Thermocyclers (PCR machines)QuarterlyThermal validation plate + probeWell-to-well uniformity ± 0.5°C
Real-time PCR instrumentsSemi-annualCalibration dye + ROI testROI within specification
Pipettes (single/multi-channel)QuarterlyGravimetric (4-point calibration)Within spec per volume range
BalancesSemi-annualCertified reference weights± 0.1 mg (analytical)
pH metersMonthly (daily use check)Buffer standards (pH 4, 7, 10)± 0.05 pH units
Freezers / refrigeratorsAnnual (sensor verification)Independent thermometer comparison± 1°C of set point
Biological safety cabinetsAnnualHEPA filter certificationISO Class 5 (Grade A)
Ultra-pure water systemSemi-annualResistivity + TOC measurement≥ 18.2 MΩ·cm, < 10 ppb TOC

Environmental Monitoring

LocationParameterFrequencyAction Limits
PCR setup roomSurface contaminationWeekly< 1 CFU/contact plate
PCR setup roomAirborne contaminationMonthly< 10 CFU/settle plate (4 hours)
DNA extraction areaSurface contaminationWeekly< 5 CFU/contact plate
DNA extraction areaAirborne contaminationMonthly< 30 CFU/settle plate (4 hours)
Amplification/post-PCR areaAmplicon contaminationWeeklyNo detectable PCR product
Clean room (reagent prep)Particulate countQuarterlyISO Class 8 or better
Storage areasTemperature mappingAnnualWithin specification throughout

Non-Conformance and CAPA

Non-Conformance SeverityDefinitionResponse TimeInvestigation Depth
CriticalPatient/result safety impact, regulatory breachImmediate (< 1 hour)Full root cause analysis
MajorQC failure, equipment malfunction affecting results< 4 hoursDetailed investigation
MinorDocumentation error, minor procedure deviation< 24 hoursSimple root cause
ObservationOpportunity for improvement< 5 business daysTrend analysis

Document Control for QC Records

Record TypeRetention PeriodStorage FormatAccess Control
QC run logs5 yearsLIMS + physical backupAll QC staff
Calibration certificatesLife of equipment + 5 yearsDocument control systemLab Manager + QC Manager
Proficiency test reports10 yearsDocument control systemQC Manager
CAPA records5 years after closureLIMSQC Manager
Audit reports5 yearsDocument control systemCEO + QC Manager
Training recordsEmployment + 3 yearsHR system + LIMSHR + QC Manager